Researchers have developed a new device that may result in more comfortable mammography for women. As per a study, standardising the pressure applied in mammography would reduce pain associated with breast compression without sacrificing image quality.

Mechanical compression of the breast in mammography often causes discomfort and pain and deters some women from mammography screening. An additional problem associated with compression is the variation that occurs when the technologist adjusts compression force to breast size, composition, skin tautness and pain tolerance. Over-compression, or unnecessarily high pressures during compression, is common in certain European countries, especially for women with small breasts. Over-compression occurs less frequently in the US, where under-compression, or extremely low applied pressure, is more common.

“This means that the breast may be almost not compressed at all, which increases the risks of image quality degradation and extra radiation dose,” said Woutjan Branderhorst, Researcher, Department of Biomedical Engineering and Physics, Academic Medical Center in Amsterdam.

Overall, adjustments in force can lead to substantial variation in the amount of pressure applied to the breast, ranging from less than three kPa to greater than 30 kPa.

Dr Branderhorst and colleagues theorised that a compression protocol based on pressure rather than force would reduce the pain and variability associated with the current force-based compression protocol. Force is the total impact of one object on another, whereas pressure is the ratio of force to the area over which it is applied.

The researchers developed a device that displays the average pressure during compression and studied its effects in a double-blinded, randomised control trial on 433 asymptomatic women scheduled for screening mammography. Three of the four compressions for each participant were standardised to a target force of 14 dekanewtons (daN). One randomly assigned compression was standardised to a target pressure of 10 kPa.

Participants scored pain on a numerical rating scale, and three experienced breast screening radiologists indicated which images required a retake. The 10 kPa pressure did not compromise radiation dose or image quality, and, on average, the women reported it to be less painful than the 14 daN force.

The study’s implications are potentially significant, Dr Branderhorst said. Pressure standardisation could help avoid a large amount of unnecessary pain and optimise radiation dose without adversely affecting image quality or the proportion of required retakes.

“Standardising the applied pressure would reduce both over- and under-compression and lead to a more reproducible imaging procedure with less pain,” Dr Branderhorst said.

The device that displays average pressure is easily added to existing mammography systems, according to Dr Branderhorst. “Essentially, what is needed is the measurement of the contact area with the breast, which then is combined with the measured applied force to determine the average pressure in the breast. A relatively small upgrade of the compression paddle is sufficient,” he added.

Further research will be needed to determine if the 10 kP a pressure is the optimal target. The researchers are also working on new methods to help mammography technologists improve compression through better positioning of the breast. Co-authors on the study are Jerry E de Groot, Mireille Broeders, Cornelis A Grimbergen and Gerard J den Heeten.